Usage : for beating of pulp according to standardized conditions (360 g).
Standards: TAPPI T200 , T205m, ISO5246-1, SCAN C25CPPA C.2
Device Description : The laboratory beater is mounted upon a stainless steel construction. The device itself is
entirely made of stainless steel as well. The container is capable of holding a volume of 23 liters , therefore 360g of
pulp can be beaten with one process .the rolls including the blade are 152mm wide and have a diameter of 190
194mm . the bed knife consists of seven 3.2mm thick stainless steel blades. The knives and bed of the beater are
encased by a tank cover in order to ensure safe handling of device.
Test Description : 360 g of pulp in a suspension is put into a beater. Now the suspension needs to be topped up
with water until the whole mixture reaches a volume of 23 liters. Three minutes after starting the device a sample
has to be taken , After that a weight of 5.5 kg has to be fixed to the lever in order to start the actual beating
process. Samples can be taken at regular intervals and the pulp can be used to be tested for example on freeness
or to form test sheets.

Accessories

Download PDF :-
Scroll to Top